Diagnostic Reference Range (DRR) in Computed Tomography (CT) Imaging Based on Clinical Indications Related to Thorax, Abdomen and Pelvic Regions
نویسندگان
چکیده
The concept of diagnostic reference range (DRR) is introduced to address the balance between a patient’s risk (radiation dose) and benefit (image quality necessary for an accurate diagnosis). R provides minimum estimated dose that covers 25–75% distribution. International Commission on Radiological Protection, ICRP (2017) recommended 75th percentile as upper d 25th lower limit DRR [1]. Dose below should compromise interpretation image above may be in excess need reviewed. Meanwhile, Diagnostic Reference Levels (DRLs) are set at distribution by guidance identify any unusual computed tomography (CT) procedures consistently exceed established DRLs.
 aim this study was establish local CT scanning based clinical indications associated thorax, abdomen, pelvic regions adult patients Department Radiology, Hospital Universiti Sains Malaysia (HUSM), Kelantan. This involved retrospective survey data underwent 5 most common related two departments, Radiology Trauma Department, HUSM. obtained human ethical approval from university committee (Protocol code: USM/JEPeM/22010025). metrics (dose length product, DLP total DLP) technical factors were collected.
 Results showed five cancer staging, coronary artery disease, pulmonary embolism, stones (CTU protocol) (KUB protocol). values varied their scanner types. Figure 1 shows boxplots representing DRRs scans Toshiba Siemens scanners. highest 4722 mGy.cm observed staging using scanner. While lowest value 40 KUB protocol also main factor contributes higher longer length.
 demonstrates descriptor values, similar anatomical region different requirement each indication. Thus, utilisation both DRL crucial minimise unnecessary patient ensure optimum produced scanning.
منابع مشابه
Thorax organ dose estimation in computed tomography based on patient CT data using Monte Carlo simulation
Background: This study presents patient specific and organ dose estimation in computed tomography (CT) imaging of thorax directly from patient CT image using Monte Carlo simulation. Patient's CT image is considered as the patient specific phantom and the best representative of patient physical index in order to calculate specific organ dose. Materials and Methods: EGSnrc /BEAMnr...
متن کاملDose Assessment in Computed Tomography Examination and Establishment of Local Diagnostic Reference Levels in Mazandaran, Iran
Background: Medical X-rays are the largest man-made source of public exposure to ionizing radiation. While the benefits of Computed Tomography (CT) are well known in accurate diagnosis, those benefits are not risk-free. CT is a device with higher patient dose in comparison with other conventional radiation procedures. Objective: This study is aimed at evaluating radiation dose to patients from ...
متن کاملthorax organ dose estimation in computed tomography based on patient ct data using monte carlo simulation
background: this study presents patient specific and organ dose estimation in computed tomography (ct) imaging of thorax directly from patient ct image using monte carlo simulation. patient's ct image is considered as the patient specific phantom and the best representative of patient physical index in order to calculate specific organ dose. materials and methods: egsnrc /beamnrc monte c...
متن کاملDiagnostic Reference Level of Computed Tomography Examinations and Need for Dose Optimization in Ondo State, Nigeria
Introduction: The present study was conducted to obtain State diagnostic reference levels (DRLs) of five routine computed tomography (CT) examinations from two CT centers in Ondo State and to identify factors responsible for dose variation and escalation in these CT centers. Material and Methods: Acquisition parameters and CT dose indices wer...
متن کاملAbsorbed dose of Fetus in SPECT-CT Imaging of Mother's Abdomen Regions using Monte Carlo method
SPECT-CT is an imaging method for diagnosing diseases and studying physiology of the body. In this type of imaging, due to the use of radiopharmaceutical for imaging (SPECT) and X-rays (CT), a relatively high dose of radiation is introduced into the patient. If the patient is pregnant, considering that the developing fetus is sensitive to any radiation in the uterus throughout the entire period...
متن کاملذخیره در منابع من