Calibration of the Greenhouse Gas Monitoring Instrument (GMI) Based on a Digital Calibration Field Network

نویسندگان

چکیده

The Greenhouse Gas Monitoring Instrument (GMI), carried by Gaofen 5 (GF5-01), and the Hyperspectral Observation Satellite (GF5-02) were successfully launched on 9 May 2018, September 7, 2021, respectively, are only passive greenhouse gas payloads in China that can regularly obtain effective detection data in-orbit at this stage. Before launch, research team out much laboratory calibration work designed an on-board system based solar radiation sources which guarantees quantitative accuracy of payload to greatest extent. In order more effectively meet high frequency requirements over whole life cycle payload, using on-track site method digital field network technology, obtained coefficient complements results. working principle GMI is quite different from a traditional imaging payload. Spatial heterodyne spectroscopy (SHS) used detect absorption spectrum gases, has large view non-imaging hyperspectral. existing fixed-site alternative methods cannot fully tasks. paper, we propose set global screening criteria characteristics design calculate coefficients non-absorption spectral channels according hyperspectral data. Based historical observation GMI, initial calculation launch was out, results four obtained: range O2 channel 1.05–1.15, mean value 1.10 standard deviation 2.72%; CO2-1 1.05–1.13, 1.09 2.64%; CH4 1.08–1.10, 1.11 2.73%; CO2-2 1.09–1.14, 1.12 2.93%. above show performance each shows no significant attenuation during period, fluctuation operation state stable.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A European Network on Digital Camera Calibration

Background The need of a camera calibration is a fundamental requirement in the context of photogrammetric data processing. For airborne sensors this calibration is typically realized under well controlled laboratory conditions. In this case, special calibration devices are used to determine the internal camera characteristics with sufficient accuracy. Using such calibration facilities, the dis...

متن کامل

EuroSDR network on Digital Camera Calibration

This paper documents the status of the actual Phase 1 of the EuroSDR project on “Digital Camera Calibration”. This project was officially installed in October 2003 during the 103 EuroSDR Science and Steering Committee Meetings. Up to now about 35 experts from industry, camera manufacturers, software developers, research and system users already joined this calibration network. Some general rema...

متن کامل

OPTIMIZATION-BASED MONITORING-SUPPORTED CALIBRATION OF A THERMAL PERFORMANCE SIMULATION MODEL

Building performance simulation is being increasingly deployed beyond the building design phase to support efficient building operation. Specifically, the predictive feature of the simulation-assisted building systems control strategy provides distinct advantages in view of building systems with high latency and inertia. Such advantages can be exploited only if model predictions can be relied u...

متن کامل

Review and Comparison of In-field Calibration Methods of Accelerometer Based on Gravity Vector

Considering the wide applications of accelerometers to determine position and attitude and due to reducing of accuracy of this sensors because of some errors, this paper discusses the calibration of accelerometers. Also because the traditional calibration methods are very time consuming, costly and need precision laboratory equipment, in-field calibration methods are recommended which are simpl...