Basic Research on Comparative Evaluation of Calculated Dose Distribution Using Monte Carlo Algorithm-equipped Treatment Planning Systems in Volumetric Modulated Arc Therapy
نویسندگان
چکیده
Purpose The purpose was to study comparative evaluation of calculated dose distribution by X-ray Voxel Monte Carlo (XVMC) for calculation in Acuros XB (AXB). commissioning and head neck volumetric modulated arc therapy (VMAT) clinical cases were compared AXB Eclipse XVMC Monaco. Methods For TrueBeam at 6 MV, we the simple rectangle, heterogeneity correction, multileaf collimator (MLC) characteristics. 15 computation times, accuracy, dose-volume histogram (DVH), 3D-γ analysis (γ 3%/2 mm). Results There no difference between values jaw field, measurement errors both within± 1%, profiles water, bone, lung equivalent slab phantoms good agreement. transmission, tongue groove effect, there a less than 10% leaf-end transmission. In cases, time half that AXB, average two calculations -1.17±2.14%, error (AXB: -0.73±0.79%, XVMC: -0.07±1.21%). DVH, max doses about 3% higher planning target volume (PTV) gross tumor (GTV), but pass rate overall 95.11±2.59%, which Conclusions Both algorithms equivalent, suggesting Monaco is verification method with high accuracy distribution.
منابع مشابه
Independent absorbed-dose calculation using the Monte Carlo algorithm in volumetric modulated arc therapy
PURPOSE To report the result of independent absorbed-dose calculations based on a Monte Carlo (MC) algorithm in volumetric modulated arc therapy (VMAT) for various treatment sites. METHODS AND MATERIALS All treatment plans were created by the superposition/convolution (SC) algorithm of SmartArc (Pinnacle V9.2, Philips). The beam information was converted into the format of the Monaco V3.3 (El...
متن کاملTreatment planning for volumetric modulated arc therapy.
PURPOSE Volumetric modulated arc therapy (VMAT) is a specific type of intensity-modulated radiation therapy (IMRT) in which the gantry speed, multileaf collimator (MLC) leaf position, and dose rate vary continuously during delivery. A treatment planning system for VMAT is presented. METHODS Arc control points are created uniformly throughout one or more arcs. An iterative least-squares algori...
متن کاملPerformance evaluation of gated volumetric modulated arc therapy
Background: Aim of this study is to evaluate the accuracy of the gated volumetric modulated arc therapy (VMAT/RapidArc) using 2D planar dosimetry, DynaLog files and COMPASS 3D dosimetry system. Materials and Methods: Pre-treatment quality assurance of 10 gated VMAT plans was verified using 2D array and COMPASS 3D dosimetry system. Advantage of COMPASS over 2D planar is that it provides the clin...
متن کاملEvaluation of volumetric modulated arc therapy for postmastectomy treatment
PURPOSE To examine the feasibility of volumetric modulated arc therapy (VMAT) for post mastectomy radiotherapy (PMRT). METHODS AND MATERIALS Fifteen PMRT patients previously treated at our clinic with helical tomotherapy (HT) were identified for the study. Planning target volumes (PTV) included the chest wall and regional lymph nodes. A systematic approach to constructing VMAT that met the cl...
متن کاملPatient-Specific Quality Assurance Using Monte Carlo Dose Calculation and Elekta Log Files for Prostate Volumetric-Modulated Arc Therapy
Log file-based methods are attracting increasing interest owing to their ability to validate volumetric-modulated arc therapy outputs with high resolution in the leaf and gantry positions and in delivered dose. Cross-validation of these methods for comparison with measurement-based methods using the ionization chamber/ArcCHECK-3DVH software (version 3.2.0) under the same conditions of treatment...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Japanese Journal of Radiological Technology
سال: 2021
ISSN: ['0369-4305', '1881-4883']
DOI: https://doi.org/10.6009/jjrt.2021_jsrt_77.8.795