Atomic Absorption Spectrometry as an Alternative to Determine the Presence of Gold Nanoparticles on or in Silica Matrix
نویسندگان
چکیده
Two different gold-silica-based nanomaterials were prepared: (i) silica-supported gold nanoparticles (AuNP/SiO2); and (ii) gold-silica core-shell (AuNP@SiO2). Three strategies for sample treatment (S), consisting in acid treatments, employed: (S1) HNO3; (S2) HNO3 + HCl; (S3) HF HCl, applying microwave oven digestion S2 S3. From three calibration curves, slope, intercept, linear correlation coefficient obtained. The accuracy of the methods was evaluated by comparing contents a determined flame atomic absorption spectrometry (FAAS) inductively coupled plasma emission (ICP-OES). Finally, amount all samples FAAS. UV-Vis spectroscopy transmission electron microscopy (TEM) used to compare material before after treatment. By comparison, application S3 allowed presence on or silica matrix be quantified.
منابع مشابه
DNA-capped mesoporous silica nanoparticles as an ion-responsive release system to determine the presence of mercury in aqueous solutions.
We have developed DNA-functionalized silica nanoparticles for the rapid, sensitive, and selective detection of mercuric ion (Hg(2+)) in aqueous solution. Two DNA strands were designed to cap the pore of dye-trapped silica nanoparticles. In the presence of ppb level Hg(2+), the two DNA strands are dehybridized to uncap the pore, releasing the dye cargo with detectable enhancements of fluorescenc...
متن کاملExtraction and Determination of Heavy Metals Using Silver Coated Magnetic Nanoparticles and Flame Atomic Absorption Spectrometry
A rapid, simple and sensitive magnetic solid phase extraction (MSPE) method was developed for the pre-concentration and determination of copper and cadmium ions. In this study, modified Fe3O4@Ag nanoparticles were synthesized and then ligand 2-isonicotinoyl hydrazine carbodithiolate (ITHCDT) was bonded to silver due to the formation of covalent bond of S-Ag. In this method, copper and cadmium i...
متن کاملUSE OF TITANIUM AND TUNGSTEN AS AN ALTERNATIVE TO GOLD IN JEWELLERY MANUFACTURING
In this study the use titanium and tungsten as alternatives to the noble metals in the jewellery industry was investigated. The degradation of titanium and tungsten were compared to that of gold, used as reference. Alternate immersion tests were performed in 3.5% sodium chloride and artificial perspiration. The metals’ abrasion resistance with respect to textile fabrics was determined. ...
متن کاملGRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY WITH MAlIlRIX MODIFICATION FOR DETERMINATION OF LEAD IN DRINKING WATER
An amoniumvanadate plus sodium molybdate mixed modifier Graphite Furnace Atomic Absorption Spectrometric method has been developed for determination of traces of lead in drinking water. The concentration of lead in the sample is obtained by direct comparison to linear working graph prepared from aqueous standards in the modifier medium, The results indicate the better performance of the pro...
متن کاملDetermination of Cadmium by Flame-Atomic Absorption Spectrometry After Preconcentration on Silica Gel Modified with Cupferron
Este trabalho descreve um método para a pré-concentração de cádmio com sílica gel modificada com cupferron e sua determinação por espectrometria de absorção atômica com chama. O efeito do pH, a quantidade do material adsorvente, a concentração de eluente, volume da amostra e quantidade do analito adsorvido foram otimizados. O efeito de interferentes foi avaliado para diversos íons mostrando que...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Journal of the Brazilian Chemical Society
سال: 2022
ISSN: ['0103-5053', '1678-4790']
DOI: https://doi.org/10.21577/0103-5053.20210158