A Varian DynaLog file‐based procedure for patient dose‐volume histogram‐based IMRT QA
نویسندگان
چکیده
منابع مشابه
A Varian DynaLog file‐based procedure for patient dose‐volume histogram‐based IMRT QA
In the present study, we describe a method based on the analysis of the dynamic MLC log files (DynaLog) generated by the controller of a Varian linear accelerator in order to perform patient-specific IMRT QA. The DynaLog files of a Varian Millennium MLC, recorded during an IMRT treatment, can be processed using a MATLAB-based code in order to generate the actual fluence for each beam and so rec...
متن کاملReproducibility in patient‐specific IMRT QA
The purpose of this study was to determine the reproducibility of patient-specific, intensity-modulated radiation therapy (IMRT) quality assurance (QA) results in a clinical setting. Six clinical patient plans were delivered to a variety of devices and analyses, including 1) radiographic film; 2) ion chamber; 3) 2D diode array delivered and analyzed in three different configurations (AP deliver...
متن کاملQA of intensity-modulated beams using dynamic MLC log files
To evaluate the utility of Dynalog file information for planar dose verification in IMRT QA, a program is developed to convert Dynalog file data to DMLC field files. For this study, five predefined fluencies are planned and delivered using Varian, Eclipse 3D planning system and 6MV photon beam of Varian, Clinac DMX linear accelerator. To measure planar dose distribution, Kodak, EDR2 films are e...
متن کاملCharacterization of cylindrical ionization chambers for patient specific IMRT QA
Proven conventional dosimetry techniques do not provide accuracy and precision in the measurement of inverse planned intensity modulated radiation therapy (IMRT) fields. Dynamic and step-and-shoot multileaf collimation (DMLC/SMLC) challenge current ionization chamber dosimetry practices. Ionization chamber performance in these fields is evaluated for three cylindrical chambers of varying volume...
متن کاملQuality Assurance of IMRT delivery systems – Varian
Introduction Intensity modulated radiation therapy (IMRT) provides the means for improved treatment for a wide variety of cancers. Piggybacked onto 3DCRT and MLC technologies, IMRT has developed quickly, but will undoubtedly see improvements in core hardware and software as well as in peripheral applications such as record and verify, gating, etc. At the core, individual treatment centers shoul...