A Validation Method for EPID In Vivo Dosimetry Algorithms
نویسندگان
چکیده
The aim of this study was to develop and apply an evaluation method for assessing the accuracy a novel 3D EPID back-projection algorithm in vivo dosimetry. Dosimetry Check (DC) 5.8 evaluated. A slab phantom homogeneously filled, or with air bone inserts, used fluence reconstruction different squared fields. VMAT plans anatomical sites were delivered on anthropomorphic phantom. Dose distributions measured radiochromic films. 2D Gamma Agreement Index (GAI) between DC film dose (3%, 3 mm) computed algorithm. GAIs films TPS also computed. within 2% all fields three slabs’ configurations. GAI 92.7% prostate, 92.9% lung, 96.6% head neck, 94.6% brain. An developed. shown be able accurately reconstruct doses anatomic sites, including lung. methodology described present can applied any
منابع مشابه
EPID in vivo Dosimetry
Introduction: The most modern radiotherapy devices are equipped with an Electronic Portal Imaging Device (EPID) system which is located on opposite side of the machine’s head. EPID system is often used to setting up the position verification during or between radiotherapy sessions. Material and Methods: Various types of dosimeters have been used to setting up ...
متن کاملEPID in vivo dosimetry in RapidArc technique.
AIM The aim of the study was to estimate the dose at the reference point applying an aSi-EPID device in the course of patient treatment. MATERIALS AND METHODS The method assumes direct proportionality between EPID signal and dose delivered to the patient reference point during the treatment session. The procedure consists of treatment plan calculation for the actual patient in the arc techniq...
متن کاملBreast in vivo dosimetry by EPID
An electronic portal imaging device (EPID) is an effective detector for in vivo transit dosimetry. In fact, it supplies two-dimensional information, does not require special efforts to be used during patient treatment, and can supply data in real time. In the present paper, a new procedure has been proposed to improve the EPID in vivo dosimetry accuracy by taking into account the patient setup ...
متن کامل3D EPID-based in vivo dosimetry for IMRT and VMAT
In this paper the various approaches of EPID-based in vivo IMRT and VMAT dose verification, and their clinical implementation, are described. It will be shown that EPID-based in vivo dosimetry plays an important role in the total chain of verification procedures in a radiotherapy department. EPID-based dosimetry, in combination with in-room imaging, is a fast and accurate tool for 3D in vivo ve...
متن کاملEvaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT
Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...