A rapid enzymatic assay for methotrexate in serum.
A kinetic enzymological assay for methotrexate in serum is described; the results are available after 1.5 hours. The lower limit of sensivity is 50 nmol/1; the day-to-day variation coefficients compare favourably with other procedures (about 7% in the 200-600 nmol/1 range). The method uses purified dihydrofolic acid reductase from bovine liver. The problem of its instability in the assay is ci...
متن کاملEnzymatic assay for methotrexate in serum and cerebrospinal fluid.
We describe an assay for methotrexate in biological fluids. The assay is based on inhibition by methotrexate of dihydrofolate reductase from Lactobacillus casei. The lower limit of sensitivity in serum is 2 X 10(-8) mol (about 10 mug) of methotrexate per liter. Within-run precision is +/- 5% (coefficient of variation) and day-to-day variation is 18%. Advantages of the assay are ease of manipula...
متن کاملA rapid, automated enzymatic fluorometric assay for determination of D-arabinitol in serum.
A rapid enzymatic fluorometric assay for measuring D-arabinitol in serum was developed using recombinant D-arabinitol dehydrogenase from Candida albicans (rArDH). rArDH was produced in Escherichia coli and purified by dye-ligand affinity chromatography. rArDH was highly specific for D-arabinitol, cross-reacting only with xylitol (4.9%) among all polyols tested. A Cobas Fara II centrifugal autoa...
متن کاملEnzymatic assay of D-mannose in serum.
We describe a new and improved enzymatic assay for determining the concentration of D-mannose in sera. Serum D-glucose is selectively converted to glucose-6 phosphate with the highly specific thermostable glucokinase (EC 2.7.1.2) from Bacillus stearothermophilus. The anionic reaction products and excess substrates are removed by a rapid and simple anion-exchange chromatography step in microcent...
متن کاملNew enzymatic assay of iron in serum.
A new enzymatic method for assaying iron in serum samples, suitable for automated analyzers, is reported. Three reagent mixtures are used: dilution buffer (pH 3.0; ascorbate), reagent 1 (pH 6.7; apoaconitase), and reagent 2 (pH 7.7; citrate, magnesium, and isocitrate dehydrogenase). Sera are diluted with dilution buffer. Fe3+ is liberated from transferrin in sera under acidic conditions, and th...