A performance comparison of the fully automated urine particle analyzer UF-5000 with UF-1000i and Gram staining in predicting bacterial growth patterns in women with uncomplicated urinary tract infections
نویسندگان
چکیده
Abstract Background The aim of this study was to compare the performance new flow cytometer UF-5000 with UF-1000i and Gram staining for determining bacterial patterns in urine samples. Methods Women who attended our clinic symptoms suggestive urinary tract infection were enrolled study. Mid-stream samples collected gram staining, analysis cultures. Bacterial classified using (none, cocci bacteria or rods/mixed growth), cocci, rods mixed growth) staining. Results Among 102 included samples, there 10 g-positive 2 bacilli, 66 g-negative rods, 24 growth. sensitivity/specificity 81.8/91.1% gram-negative 23.5/96.9% cocci/mixed. 80.0/88.2% negative 70.0/86.5% gram-positive cocci. Conclusions demonstrated good sensitivity specificity Gram-negative bacilli an improved detecting Gram-positive compared UF-1000i.
منابع مشابه
[Modifying cutoff values of the UF 1000i(®) automated urine analyzer to predict urinary tract infection].
The fundamental advantage of using automated urine analyzers lies in the current grouping of laboratories as technical divisions. The UF1000i(®) Biomérieux is a 2(nd) generation automated urine analyzer which uses fluorescence flow cytometry to count the particles and bacteria present in urine. It also provides a prediction indicator for the urinary tract infection, based on the leukocyte and b...
متن کاملEvaluation of the Sysmex UF-1000i for the diagnosis of urinary tract infection.
Diagnosis of urinary tract infection (UTI) is primarily done by microbiologic culture, which is time-consuming and can produce false-positives and false-negatives. Flow cytometry allows for rapid screening of many samples and eliminates culturing. We analyzed the Sysmex UF-1000i (TOA Medical Electronics, Kobe, Japan) for accuracy in identifying RBCs and WBCs, casts, bacteria, and epithelia. We ...
متن کاملthe past hospitalization and its association with suicide attempts and ideation in patients with mdd and comparison with bmd (depressed type) group
چکیده ندارد.
Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures.
UF-100 flow cytometer and urine strip results were cross-interpreted to predict culture outcomes. The best negative predictive value was obtained with bacteria at > or =1,000/microl, white blood cells at > or =20/microl, or leukocyte esterase positivity. Nine of 24 false negatives were clinically significant. Thus, UF-100 and urine strip results do not accurately predict the outcome of cultures.
متن کاملEvaluation of the Sysmex UF-100 automated urinalysis analyzer.
Urinalysis is a high-volume procedure that currently requires significant labor to examine microscopic sediment. We evaluated the Sysmex UF-100 automated urinalysis analyzer for performing this task. Instrument accuracy was assessed by comparing continuous counts of microscopic elements from the UF-100 with ranges of cells (per low-power field or high-power field) from manual microscopy perform...